SPELEC is a tool for spectroelectrochemical measurements. It comprises a spectrometer (UV-VIS wavelength range: 200-900 nm), a bipotentiostat/galvanostat, and a lightsource in one box.
It also comes with specialized spectroelectrochemical software that enables the synchronization of optical and electrochemical studies.
Tech Specs
Source: Metrohm USA Inc.
. | . |
---|---|
Operating modes | Bipotentiostat; Potentiostat; Galvanostat |
PC interface | USB |
Power | 12 V DC |
Potential range in volt | ±4 V |
Maximum current in Ampère | ±40 mA |
Number of current ranges | 8 |
Number of current ranges remarks | ±1 nA to ±10 mA |
Number of potential ranges (galvanostat) | 2 |
Number of potential ranges remarks (galvanostat) | ±100 mV to ±1 V |
Multichannel instrument | NO |
Maximum number of channels | 1 |
Maximum number of WE sharing AUX and REF | 2 |
Potential accuracy | ±0.2% |
Current accuracy | ≤0.5% of current range at 100 nA to 10 mA |
Potential resolution | 1 mV |
Current resolution | 0.1% of current output range |
Measured potential resolution | 0.012% of potential range |
Measured current resolution | 0.025% of current range |
Spectrometer (SPM) Detector | Linear silicon CCD array |
SPM Pixels | 2048 |
SPM Pixel size | 14 µm x 200 µm |
SPM Pixel well depth | 62,500 electrons |
SPM Wavelength range | 200-900 nm |
SPM Optical resolution | 0.3+10.0 nm FWHM |
SPM Signal-to-noise ratio | 250:1 (at full signal) |
SPM A/D resolution | 16 bit |
SPM Dark noise | 50 RMS counts |
SPM Dynamic range | 8.5 x 10 rised to 7 (system) ; 1300:1 for a single acquisition |
SPM Integration time | 1 ms to 65 seconds |
SPM Stray light | ≤ 0.05% at 600 nm ; < 0.10% at 435 nm |
Light Source (LS) wavelength range | 215-400 nm (deuterium) ; 360-2500 nm (tungsten halogen) |
LS Stability | 1.0% peak-to-peak (over 4 hours) after 30-minute warm-up |
LS Time to stable output | 10 minutes (deuterium) ; 1 minute (tungsten halogen) |
LS Ignition delay | <2.0 seconds (delay for cold start-up may be longer) |
LS Bulb life | >1,000 hours @ 240 nm (time); <50% @ 240 nm (decrease of intensity) ; Continuous operation (testing conditions) |
LS Fiber optic connector | SMA 905 |
Dimensions | 24 x 11 x 25 |
Weight of the instrument in kg | 1.95 |
Velmex, Inc. is a premier North American manufacturer of standard and custom linear and rotary mo……...
The VFG Shotcrete Penetrometer, a specialized force testing kit, is tailored for evaluating spray……...
Here at PEM we specialise in manufacturing hand crafted stainless steel products to your individu……...
Decatur Plastic Products (DPP) provides high quality injection molding, flocking, assembly and fi……...
Kati Oy is a contractor company specializing in ore exploration and rock construction services, w……...
Three types of Planar Cell Test Manifold Kits are available from fuelcellmaterials to handle larg……...
SupplyOne is the leader in packaging solutions. We build a packaging program that simplifies your process, guarantees savings, and increases efficiency. Let us do the work while you call the shots……
Nanophase develops, manufactures and sells an integrated family of nanomaterial technologies. Nanophase produces engineered nanomaterial products for use in a variety of markets includingSurface Fi……
M.R. MOLD AND ENGINEERING provides precision tooling for the rubber and plastic industries. We specialize in liquid silicone rubber injection, gum stock rubber injection and plastic injection molds……