Optilab Refractive Index Detector for HPLC/GPC

上海仪舶实验室自动化 25-01-23 17:12:09

Optilab Refractive Index Detector for HPLC/GPC

The Optilab® differential refractive index (dRI) detector may be used in conjunction with any HPLC, FPLC, GPC or FFF separation system, plus a Wyatt DAWN® or miniDAWN® multi-angle light scattering (MALS) detector, for the absolute determination of the molar mass of all types of macromolecules in solution. While its sensitivity matches the very best HPLC RI detectors, its dynamic range exceeds that of any comparable instrument by 10-100x, with no range settings - meaning you can characterize very large and very small peaks, simultaneously.

In addition to dRI, the Optilab also measures the absolute refractive index (aRI) of a solution, important for characterizing solvents. The high-concentration (HC) model provides an extended measurement range suitable for high-concentration samples (up to 180 mg/mL of protein) or gradient-elution chromatography such as ion-exchange.

Features

dRI noise: 7.5x10-10 RIU (HC model: 1.5x10-9 RIU)

dRI range: ±4.7x10-3 RIU (HC model: -2.6x10-3 to +34x10-3 RIU)

Dynamic range: 12,000,000:1 (HC model: 23,000,000:1)

Absolute RI sensitivity: ±0.002 RIU

Absolute RI range: 1.2 – 1.8 RIU

Temperature control: heated/cooled 4 - 65°C, including active stabilization—even at room temperature

Wavelength: 660 nm, matched to MALS instruments; other wavelengths available

Touch-panel display for immediate access to data, diagnostics and settings

Applications

Universal online concentration measurement for chromatography, especially SEC-MALS and FFF-MALS.

Analysis of conjugated proteins, co-polymers and nanoparticle-encapsulated drugs/proteins/DNA/RNA when combined with a MALS and UV detector

Offline measurement of second virial coefficient A2 (B22)

Offline measurement of the specific refractive index increment dn/dc, at the same wavelength as the MALS laser (critical for accurate macromolecular characterization)

Ion-exchange chromatography with MALS (IEX-MALS) – measure dRI over the entire salt gradient

Composition-gradient MALS (CG-MALS) – measure protein-protein interactions at high concentration

[Optilab Refractive Index Detector for HPLC/GPC]相关推荐

美国New York University (NYU)

自 1831 年成立以来,纽约大学一直是高等教育的创新者,面向新兴中产阶级,拥抱城市特色和专业重点,并推……...

LEDVANCE GmbH

LEDVANCE GmbH

LEDVANCE offers a wide range of LED luminaires, intelligent and connected Smart Home solutions, m……...

PARDAM NANO4FIBERS

PARDAM NANO4FIBERS

PARDAM NANO4FIBERS sro . develops high-quality nanofibrous materials NnF CERAM ® and nanofibr……...

今日仪器
  1. Item North America

    Item North America

    Item North America is a premier provider of precision ball screws for use as a drive mechanism in linear slides, multi-axis motion and complex material handling devices. item ball screws feature hi……

    代理品牌 2025-09-04

  2. Streaming Answers: SurPASS™ 3

    Streaming Answers: SurPASS™ 3

    SurPASS™ 3 is capable of measuring the zeta potential over an extensive range of materials and allows the analysis of varied surface properties and their changes. The zeta potential explains ……

    仪器产品 2025-09-04

  3. AHP Materials

    AHP Materials

    AHP Materials is a specialty materials company focused on the manufacturing and selling of high purity specialty metals and compounds. AHP concentrates mainly on Antimony, Cadmium Sulfide, Telluriu……

    代理品牌 2025-09-04

返回顶部小火箭