[db:国别翻译]加拿大实验室
Micom Laboratories Inc. 是一家获得 ISO/IEC 17025 (2017) A2LA 认证的第三方独立实验室,提供产品和……...

The FOX 50 Heat Flow Meter from TA Instruments is a microprocessor-based instrument for testing materials in the conductivity range of 0.1 W/mK to 10 W/mK.
The FOX 50 Heat Flow Meter is an easy-to-use, accurate instrument for measuring thermal conductivity and delivers fast results in a compact footprint. The instrument is configured with the same high performance features and proprietary technologies of the larger FOX systems including thin film heat flux transducers, responsive temperature control, digital thickness measurements as well as an integrated contact-resistance correction.
Covering a broad range of temperatures, the FOX 50 is a perfect choice for measurements of medium-conductivity materials such as plastics, glasses, concrete, composites, ceramics and more.
Key Features
Conforms to ASTM C518 and ISO 8301
Compact size and cost-effective system for thermal conductivity measurements
Solid state cooling/heating for precise temperature control
Proprietary thin film heat flux transducers for the most representative sample heat flow measurement
Optical encoder for the most accurate digital measurement of sample thickness
Robust WinTherm-50 software for improved testing functionality including heat capacity
Pyrex reference standards for calibration and verification. Accurate results can be produced for years between calibrations.
Interfacial resistance correction (two-thickness method)
Automatic sample feeder for high-throughput analyses
Optional software for specific heat measurements of solids and liquids
Optional liquid cell for testing of fluids
Specifications
| . | . |
|---|---|
| Sample Diameter | 51 mm (2") |
| Temperature Range |
-10 °C – 110 °C FOX 50 0 °C – 190 °C FOX 50 VHS |
| Temperature Resolution | ± 0.01 °C |
| Thermal Conductivity Range | 0.1 to 10 W/(mK)(0.633 to 60.3 BTU in/hr ft2 °F) |
| Thermal Resistance Range | 0.003 to 0.05 m2K/W |
| Accuracy, Standard | ± 3% |
| Accuracy, (VHS) | ± 4% |
| Reproducibility | ±2% |
| Heat Flux Transducers Area | 25 x 25 mm (1 sq. in.) |
| Instrument Dimensions | 250 mm width, 170 mm depth, 360 mm height |
| Instrument Weight | 11 Kg |
| Power Requirement | 115 V or 220 V, 50/60 Hz |
Micom Laboratories Inc. 是一家获得 ISO/IEC 17025 (2017) A2LA 认证的第三方独立实验室,提供产品和……...
Christy Metals, Inc. is an Illinois-based private corporation supplying industry with the finest ……...
GEI Inc. is dedicated to the design and manufacturing of LCD, LED, RELAYS, and TRANSFORMERS for N……...
Engineering and manufacturing custom ceramic and ferrite components for over 45 years. Since 197……...
Swabian Instruments is a test and measurement company headquartered in Stuttgart, Germany. We dev……...
Leti is an institute of CEA, a French research-and-technology organization with activities in ene……...
Item North America is a premier provider of precision ball screws for use as a drive mechanism in linear slides, multi-axis motion and complex material handling devices. item ball screws feature hi……
SurPASS™ 3 is capable of measuring the zeta potential over an extensive range of materials and allows the analysis of varied surface properties and their changes. The zeta potential explains ……
AHP Materials is a specialty materials company focused on the manufacturing and selling of high purity specialty metals and compounds. AHP concentrates mainly on Antimony, Cadmium Sulfide, Telluriu……