Angstrom Advanced Offers Portable X-ray Flaw Detector
The AXFG series detector available from Angstrom Advanced is fitted with a rippled ceramic x-ray ……...
The DIL805 Series from TA Instruments has four models - DIL 805L and DIL 805A are quenching dilatometers, DIL 805A/D is a quenching dilatometer with the ability to deform the specimen by compression, and DIL 805A/D/T can also stretch it in tension mode.
All four instruments are totally automated, self-contained units used to measure dimensional changes under challenging conditions of controlled cooling and heating.
In the quenching mode the sample, hollow or solid, is inductively heated to a temperature plateau and is then cooled at a user-defined (exponential or linear) cooling rate. The phase transformation occurring in the nonstop cooling process or in the isothermal dwell, with or without compression/tensile stress, is specified by the measured change in length.
A range of isothermal or cooling curves signifies an isothermal time-temperature-transformation (TTT) diagram or a continuous-cooling-transformation (CCT) diagram respectively. The DIL 805A specifies the benchmark for establishing these dimensional changes and phase transitions. Functioning from -160 °C up to 1500 °C (in two diverse furnace configurations) with heating rates of up to 4000 °C/s and cooling rates of 2500 °C/s, can closely mimic the material response for any heat treatment process or production.
The DIL 805A/D, on top of the quenching mode, is well-known for its capability to deform the specimen with controlled deformation rates of 0.01 to 200 mm/s. Used to improve steel processes like hot or cold rolling, the DIL 805A/D allows to develop time-temperature-transformation diagrams after deformation (DTTT) and is also used to test creep and relaxation processes.
The DIL 805A/D/T additionally extends the capabilities to alternate tensile and compressive loading to emulate mill processing. Furthermore, tensile loading to fracture lends extra information about material’s final performance and allows creation of true-stress vs true-strain or stress/strain cycling plots.
Specifications
DIL 805L | DIL 805A | DIL 805A/D | DIL 805A/D/T | |
---|---|---|---|---|
Temperature Range(sample dependent) | -150 °C to 1300 °C50 °C to 1500 °C | -150 °C to 1300 °C50 °C to 1500 °C | 50 °C to 1500 °C | 50 °C to 1500 °C |
Heating principle | inductive | inductive | inductive | inductive |
Sample material | Electrically conductive hollow or solid body | Electrically conductive hollow or solid body | Electrically conductive solid body | Electrically conductive solid body |
Sample Geometry | OD 4 mmLength 10 mm | OD 4 mmLength 10 mm | OD 5 mmLength 10 mm | OD 5 mmLength 10 mm |
Length Resolution | 50 nm | 50 nm | 50 nm | 50 nm |
Temperature Resolution | 0.05 °C | 0.05 °C | 0.05 °C | 0.05 °C |
Atmosphere | inert gas, vacuum, air | inert gas, vacuum, air | inert gas, vacuum, air | inert gas, vacuum, air |
Heating rateCooling rate | 2000 °C/sec2500 °C/sec | 4000 °C/s2500 °C/sec | 100 °C/s100 °C/sec | 100 °C/s100 °C/sec |
Deformation force | up to 20.0 kN | up to 8.0 kN | ||
Deformation rate | 0.01 – 200 mm/s | 0.01 – 20 mm/s | ||
Strain rate φ | 0.001 – 20.0 s-1 | 0.001 – 20.0 s-1 | ||
True strain φ | 0.05 – 1.2 | 0.05 – 1.2 | ||
Minimum Gage Length | 3 mm | 3 mm | ||
Number of deformation steps | any number | any number | ||
Pause between deformation steps | 60 msec | 60 msec |
The AXFG series detector available from Angstrom Advanced is fitted with a rippled ceramic x-ray ……...
Dip Seal Plastics manufactures a variety of protective strippable plastic coatings. Dip Seal stri……...
The WIRTGEN GROUP is an internationally operating group of companies in the construction equipmen……...
RMI Pressure System (a member of Armstrong Fluid Technology) is a global leader with an enviable ……...
We are committed to the high standards of durability and consistency required by the architectura……...
HORIBA Scientific is the new global team created to better meet customers’ present and futu……...
A/D Fire Protection Systems' quality goal is to cost effectively manufacture passive fire protection materials that meet or exceed regulatory requirements and the expectations of our customers……
Allegheny Bradford Corporation is a Pennsylvania corporation with three operating divisions—Allegheny Bradford, Top Line Process Equipment Company, and Allegheny Surface Technology—whic……
Video credit: Thermo Fisher Scientific Using the Thermo Scientific™MarqMetrix™All-in-One Process Raman Analyzer, organizations have the ability to perform instantaneous, non-invasive a……