RHK’s VT Beetle is based on the field-proven “Johnny Walker” Beetle™ design. It is both thermally and mechanically stable compared to the many SPM designs with varying temperatures that are available on the market today.
The Beetle STM with broad VT operation ranging between 25 K (LHe) and 1500 K, has provided superior quality low current imaging for more than 10 years. Flexible functioning modes along with RHK Technology’s high bandwidth control and low noise electronics provide users with an exceptional STM system package.
The company’s patented tip exchange system enables simple tip exchange with either an optional tip heating module or in-situ tip conditioning on VT systems.
Specifications
The specifications of the Beetle UHV VT STM are as follows:
Beetle design scans faster than sample scanners due to fewer restrictions on resonant frequency
High speed scanning: 3 second/frame
Minimum resolution of 0.5Å (X, Y) and 0.1Å (Z)
Low current imaging: < 1 pA at 5 kHz bandwidth
Sample temperatures from 25 K to >1500 K on SPM sample stage and on manipulator stages in preparation/analysis chambers
Suitable for spectroscopy and extended observation of a specific feature
Heating samples have no impact on Scan Tube, preventing the risk of affecting its thermal coefficient and causing drift
Coarse XY positioning no less than 5 mm x 5 mm
Sample size 10 mm diameter
Sample holders also available for cleaving; conveying RF signal to sample; and heating in reactive gases via Quartz Bulb or Ceramic Button
Thermal drift <1 Å/minute
Inherent mechanical and thermal stability due to small mechanical and thermal loop, inherently stiff and compact Beetle design, and symmetrical geometry
Sample temperatures: from 25 K (LHe) to >1500 K; or 100 K (LN2) to >1500 K
Scan range of at least 5 µ x 5 µ XY
Beetle sample holder uniquely provides true and accurate sample temperature measurement via thermocouple in direct contact with sample itself.
Sample temperature while imaging: 750 K for AFM, 1000 K for STM
Six electrical contacts to sample holder (two for heating, two for temperature, two for auxiliary use such as BEEM). A single holder conveniently supports all heating and cooling modes – radiative, e-beam, resistive/DC; LN2 and LHe
Red Pine Exploration has a long tradition of gold exploration and production in Canada, including……...
Laser Materials Corporation has been in business for over 10 years growing large size Nd:YAG crys……...
Atomica Corp. unleashes the power of Micro Electro-Mechanical Systems (MEMS) to help solve the gr……...
Alucoat manufactures industrial grade aluminium honeycomb, under the brand name ALUNID honeycomb.……...
Beckman Coulter Life Sciences is dedicated to empowering discovery and scientific breakthroughs. ……...
PrepQuip is a South African based company, supplying equipment to the mining, metallurgical and m……...
A/D Fire Protection Systems' quality goal is to cost effectively manufacture passive fire protection materials that meet or exceed regulatory requirements and the expectations of our customers……
Allegheny Bradford Corporation is a Pennsylvania corporation with three operating divisions—Allegheny Bradford, Top Line Process Equipment Company, and Allegheny Surface Technology—whic……
Video credit: Thermo Fisher Scientific Using the Thermo Scientific™MarqMetrix™All-in-One Process Raman Analyzer, organizations have the ability to perform instantaneous, non-invasive a……