GaGe
GaGe, a Vitrek brand, established in 1987, is a US-based producer in high-speed data acquisition ……...
The Airnet II 4-Channel particle sensor simplifies cleanroom monitoring in a cost-effective manner. With its compact design, it delivers exceptional performance and has the capability to transmit data. It also complies with the ISO 21501-4 and ISO 14644-1:2015 standards.
The sensor is easy to install, offering various power options, including distributed power from an in-house system, local power sources, or Power over Ethernet (PoE).
It also supports various communication methods, including Ethernet for interfacing with Pharmaceutical Net, Facility Net, or FacilityPro® software, OPC communications, Modbus communications, or an optional 4-20 mA output.
The sensor ensures data integrity by continuing to collect data even during network disruptions, thanks to its data queue feature. Furthermore, it includes a Dynamic Flow Sensing system to guarantee consistent flow conditions and vacuum system functionality, alerting users to any 15% change in flow conditions.
Benefits
A laser diode (LD) drastically reduces the need for maintenance and extends product lifecycle
A low-cost solution for multipoint monitoring
A small footprint and flexible mounting options make it easy to install in cleanrooms and mini-environments
Allows for immediate reaction to particle contamination events
Automatic laser shutdown reduces laser failures
Data queue maintains data integrity when communication is lost
Interfaces with Facility Net, Pharmaceutical Net and FacilityPro Software for comprehensive management of cleanroom conditions
Optional 4-20 mA output for integration with existing systems to help you understand your environment
Proven technology provides reliable and accurate data
Features
0.1 and 1.0 CFM flow rate
0.2 and 10.0 µm size range
4 channels
Chemical-resistant polycarbonate (PC) enclosure
Includes system validation documentation
Interfaces with object linking and embedding (OLE) for process control (OPC) communications, Modbus communications and optional 4-20 mA output
ISO14644-1:2015 Compliant
Low sample point cost
Small enough for use in remote locations
Applications
Cleanroom monitoring
Dedicated monitoring of critical locations
Isolator monitoring
Multi-location monitoring
Statistical process control
Trend analysis
Specifications
Source: Particle Measuring Systems
201-4 | 301-4 | 310-4 | 501-4 | 510-4 | |
---|---|---|---|---|---|
Size Range (μm) | 0.2, 0.3,0.5, 1.0 | 0.3, 0.5,1.0, 5.0 | 0.3, 0.5,1.0, 5.0 | 0.5, 1.0,5.0, 10.0 | 0.5, 1.0,5.0, 10.0 |
Flow Rate | 0.1 CFM(2.8 LPM) | 0.1 CFM(2.8 LPM) | 1.0 CFM(28.3 LPM) | 0.1 CFM(2.8 LPM) | 1.0 CFM(28.3 LPM) |
Counting Efficiency | 50% ± 20% for most sensitive channel. Meets ISO 21501-4100% ± 10% at 1.5 to 2.0 times channel one size. Meets ISO 21501-4 | ||||
Zero Count | ≤ 70.7 counts/m3 | ≤ 70.7 counts/m3 | ≤ 7.07 counts/m3 | ≤ 70.7 counts/m3 | ≤ 7.07 counts/m3 |
Maximum Concentration1 | 5,057,310 /ft3 | 4,862,798 /ft3 | 702,404 /ft3 | 7,437,220 /ft3 | 890,371 /ft3 |
Laser Source | Diode | ||||
Laser Classification | Class 1 per EN60825 (Internally, a Class IIIB laser is used, per EN60825) | ||||
Exterior Surface | Polycarbonate | ||||
Dimensions(H x W x L) | 3.8 x 3.6 x 5.3 in (9.6 x 8.9 x 13.5 cm) | ||||
Weight | 1.6 lb (0.73 kg) | ||||
Sample Probe or Tubing | 1/4” ID | ||||
Flow System | External vacuum 1/4” connectionAutomatic laser shutoff and alarm on 15% flow variation | ||||
Vacuum Source | > 12 in Hg required | ||||
Power | Power over Ethernet (PoE) via PoE router (48 VDC) or PoE power injectorOptional 24 VDC (0.5 A) power input | ||||
Communication Connectors | Ethernet (Particle Measuring Systems proprietary protocol, OPC, Modbus TCP)Optional 4-20 mA (5 output channels: 4 particle data, 1 instrument status)RS-232 (configuration and diagnostic only) | ||||
Status Indicators | Programmable status (two-color LED), Activity (one-color LED) | ||||
Calibration | Calibration materials used are traceable to the National Institute for Standards and Technology (NIST)and meet ISO 21501-4 requirements | ||||
Environment | Temperature: 39 – 95 °F (4 – 35 °C)Relative humidity: 5 – 95%, non-condensing | ||||
Complies With | EU RoHS, ISO 21501-4 |
1Greater than 90% accuracy (less than 10% coincidence loss)at maximum recommended concentration.
GaGe, a Vitrek brand, established in 1987, is a US-based producer in high-speed data acquisition ……...
The GeoSpec is the industry standard NMR core analyzer, with installations in nearly every key oi……...
Yissum is the technology transfer company of the Hebrew University of Jerusalem (HU). It is respo……...
Ram Specialty Fabrications is a full Sheet Metal / Fabrication House in Naugatuck, CT. We service……...
Founded in 2004, Magritek is the global leader in manufacturing cryogen-free benchtop Nuclear Mag……...
NSRW 公司专营铍铜、铬铜、铜合金以及电阻焊设备及耗材。我们设计、制造和翻新设备和工具。此外,我们……...
Item North America is a premier provider of precision ball screws for use as a drive mechanism in linear slides, multi-axis motion and complex material handling devices. item ball screws feature hi……
SurPASS™ 3 is capable of measuring the zeta potential over an extensive range of materials and allows the analysis of varied surface properties and their changes. The zeta potential explains ……
AHP Materials is a specialty materials company focused on the manufacturing and selling of high purity specialty metals and compounds. AHP concentrates mainly on Antimony, Cadmium Sulfide, Telluriu……